In Reply:--Noel points out that the catheters used in our study [1] retrieved less than half of the venous air embolus and that a continuous column of frothy blood extending from the superior vena cava into the pulmonary vasculature was noted on necropsy. He indicates that, if the catheters were positioned in the pulmonary artery, a larger portion of the air embolus may have been extracted, and more animals would have survived. The rationale for positioning multiorifice catheters at the superior vena cava-right atrial (SVC-RA) junction was based on previous in vitro studies that concluded that an air lock occurs in the right atrium. Those evaluations documented that more than 80% of an air embolus can be removed by a multiorifice catheter located at the SVC-RA junction. [2,3] The in vivo animal studies have confirmed that the majority of the air can be removed by a catheter located at the SVC-RA junction. [4-7] However, the design of those studies may have contributed to those findings. Those studies injected air into the internal jugular or other large central vein. This may cause streaming of the air and a right heart air lock. Our study design attempted to mimic a venous air embolism occurring from a dural sinus during a seated procedure.
Because of the results of our investigation, we postulate that there may be a fundamental difference in the blood fluid interface presented to the catheters when air is injected into a major vessel as opposed to a dural sinus. The air injected into the dural sinus may undergo considerable fractionation, as opposed to air injected into the central circulation. Is the frothy mixture less capable of creating a right heart air lock and thus less capable of aspiration? If so, it would explain the relatively low percentage of air aspirated compared to the other animal study designs. Based on the low volume of aspiration and the necropsy findings, should we reevaluate the location of the catheter for air aspiration? Would a high-volume aspiration catheter in the pulmonary vasculature be more effective than one at the SVC-RA border? Questions remain to be answered in future investigations.
Regarding the design of the catheter, Noel is correct that the electrocardiogram electrode design of the JX-318 catheter (Arrow International, Reading, PA) is probably better. The design of the JX-318 catheter for aspiration of air, however, appears to be inferior and has never been evaluated in any study of air aspiration.
Paul D. Mongan, M.D., Jon A. Hinman, D.V.M., M.D., Anesthesia and Operative Service, Department of Surgery, Brooke Army Medical Center, San Antonio, Texas 78234-6200.
REFERENCES
Mongan PD, Hinman JA: Evaluation of a double-lumen multiorifice catheter for resuscitation of swine from lethal venous air embolism. ANESTHESIOLOGY 1995; 83:1104-11.
Bunegin L, Albin MS, Helsel PE, Hoffman A, Hung TK: Positioning the right atrial catheter: A model for reappraisal. ANESTHESIOLOGY 1981; 55:343-8.
Hanna PG, Gravenstein N, Pashayan AG: In vitro comparison of central venous catheters for aspiration of venous air embolism: Effect of catheter type, catheter tip position, and cardiac inclination. J Clin Anesth 1991; 3:290-4.
Colley PS, Artru AA: Bunegin-Albin catheter improves air retrieval and resuscitation from lethal venous air embolism in dogs. Anesth Analg 1987; 66:991-4.
Colley PS, Artru AA: Bunegin-Albin catheter improves air retrieval and resuscitation from lethal venous air embolism in upright dogs. Anesth Analg 1989; 68:298-301.
Artru AA: Modification of a new catheter for air retrieval and resuscitation from lethal venous air embolism: Effect of nitrous oxide on air retrieval. Anesth Analg 1992; 75:226-31.
Artru AA: Venous air embolism in prone dogs positioned with the abdomen hanging freely: Percentage of gas retrieved and success rate of resuscitation. Anesth Analg 1992; 75:715-9.