Endotracheal Tube Obstruction after Orogastric Tube Placement 
Author Notes
  • (Pousman) Fellow, Trauma Anesthesia and Critical Care.
  • (Koch) Assistant Professor of Anesthesiology and Internal Medicine.
  • Received from the Department of Anesthesiology, The University of Texas Medical School, Houston, Texas. Submitted for publication April 18, 1997. Accepted for publication June 12, 1997.
  • Address reprint requests to Dr. Pousman: Department of Anesthesiology, The University of Texas Medical School, 6431 Fannin, Suite 5.020, Houston, Texas 77030. Address electronic mail to: bpousman@anesl.med.uth.tmc.edu.
Article Information
Case Reports
Case Reports   |   November 1997
Endotracheal Tube Obstruction after Orogastric Tube Placement 
Anesthesiology 11 1997, Vol.87, 1247-1248. doi:
Anesthesiology 11 1997, Vol.87, 1247-1248. doi:
The placement of a gastric suction tube is a common practice in the operative suite and in the intensive care unit (ICU) and rarely results in a complication. Reports of pulmonary complications associated with gastric tube placement center around their insertion into the tracheobronchial tree. [1–3 ] Cases of gastric tubes entangled with endotracheal tubes (ETT), producing restriction of gas flow, are rare. [4 ] We present a patient with airway compromise resulting from an orogastric tube knotted around the ETT and requiring emergent reintubation.
Case Report 
A 46-yr-old woman was transferred to our facility for treatment after a self-inflicted gunshot wound to the head. At the scene, her Glasgow Coma Score was 7, and her trachea was intubated orally. Physical examination revealed right parietal entrance and left parietal exit wounds with bilateral hemotympanum. In the ICU, an orogastric tube (OGT) was placed to relieve gastric distention. The oral route was chosen because of concern for a possible basilar skull fracture. After placement of the OGT, the epigastrium was auscultated with no sounds heard with air insufflation. The OGT was partially withdrawn and reinserted. The OGT was then placed on wall suction with no output noted and again, repositioned without removal. Resistance was noted whenever the OGT was withdrawn or advanced with concomitant movement of the ETT. At this time, the high pressure alarm on the ventilator sounded revealing peak airway pressures of 50 cmH20. Arterial pulse oximeter alarmed and oxygen saturation decreased to 88%. An attempt to pass a suction catheter through the ETT was unsuccessful. The patient was disconnected from the ventilator, and bag ventilation was attempted, but no air movement was noted. While extubating the patient, the OGT also was removed. The patient was easily reintubated using direct laryngoscopy, which revealed no abnormalities of the vocal cords or laryngeal structures. She was then placed back on mechanical ventilation with normal airway pressures, and bilateral chest expansion was noted. After clinical stabilization, the ETT was inspected and found to have the OGT knotted around the distal end, thus constricting the lumen (Figure 1).
Figure 1. Endotracheal tube with knotted orogastric tube.
Figure 1. Endotracheal tube with knotted orogastric tube.
Figure 1. Endotracheal tube with knotted orogastric tube.
×
Discussion 
Insertion of gastric tubes into the tracheobronchial tree has been reported and may cause pulmonary complications, especially if left unrecognized. Complications include infusion of enteral feeds, with subsequent infection, direct mechanical trauma to the airway, and ventilatory compromise. [1–3 ]
Other reports of ventilatory compromise after insertion of gastric tubes focus on the reduction of delivered tidal volumes. This reduction in tidal volume may be caused by a gastric tube that has been placed in the trachea past the ETT cuff with resulting decreased competency, allowing volume to escape. [5 ] Further, negative pressure applied to a misplaced nasogastric tube also has been implicated as a cause of decreased ventilation. [5,6 ]
We have presented a case of a gastric suction tube impeding ventilation caused by knotting around the ETT. We believe that multiple manipulations of the OGT may have led to knotting. Recognition of entanglement of the OGT with the ETT was diagnosed by the concomitant tube motion when traction was applied to the OGT. The inability to pass a suction catheter through the ETT further confirmed ETT compromise and the need for intervention. This complication, ETT-OGT entanglement, needs to be recognized and managed expeditiously. Possible treatment options include the pas-sage of an tube exchanger catheter through a patent lumen, removing the ETT and sliding a new one over the exchanger catheter, using direct laryngoscopy to cut the tube off the ETT, [4 ] or, as in our case, extubation and reintubation.
References 
References 
Eldar S, Meguid MM: Pneumothorax following attempted nasogastric intubation for nutritional support. JPEN J Parenteral Enteral Nutrition 1984; 8:450-2.
Torrington KG, Bowman M: Fatal hydrothorax and empyema complicating a malpositioned nasogastric tube. Chest 1987; 79:240-2.
el-Gamel A, Watson DC: Tracheobronchial intubation of the right pleural space: A rare complication of nasogastric intubation with a polyvinylchloride tube-A case study. Heart Lung 1993; 22:224-5.
Bautista EM: Complications of nasogastric tube insertion (Letter). Chest 1988; 93:1119-20.
Zwagil LW, Metzerott KO: Mechanical ventilation compromised by misplacement of nasogastric tube: report of an unusual incident. Respir Care 1973; 18:60-1.
Wood G, Milne B, Spjeda V, et al: Ventilatory failure due to an improperly placed nasogastric tube. Can J Anaesth 1990; 37:587-8.
Figure 1. Endotracheal tube with knotted orogastric tube.
Figure 1. Endotracheal tube with knotted orogastric tube.
Figure 1. Endotracheal tube with knotted orogastric tube.
×