Pain Medicine  |   October 2009
Mortality Associated with Implantation and Management of Intrathecal Opioid Drug Infusion Systems to Treat Noncancer Pain
Author Affiliations & Notes
  • Robert J. Coffey, M.D.
    *
  • Mary L. Owens, M.D.
  • Steven K. Broste, M.S.
  • Michel Y. Dubois, M.D.
    §
  • F Michael Ferrante, M.D.
    ||
  • David M. Schultz, M.D.
    #
  • Lisa J. Stearns, M.D.
    **
  • Michael S. Turner, M.D.
    ††
  • *Medical Advisor, Neuromodulation Clinical and Emerging Therapies, † Medical Safety Director, Neuromodulation Regulatory Affairs, ‡ Director, Neuromodulation Biostatistics, Medtronic Inc., Minneapolis, Minnesota; § Professor of Anesthesiology, Director, Education and Research, New York University Pain Program, New York University Langone Medical Center, New York, New York; || Professor of Clinical Anesthesiology, Director of Pain Medicine, David Geffen School of Medicine, University of California, Los Angeles, California; # Medical Director, MAPS Medical Pain Clinics, Minneapolis, Minnesota; ** Medical Director, The Center for Pain and Supportive Care, Scottsdale, Arizona; †† Clinical Assistant Professor of Surgery, Indiana University School of Medicine, Indianapolis, Indiana, and Adjunct Assistant Professor of Pharmacy, Purdue University, West Lafayette, Indiana.
Article Information
Pain Medicine / Pain Medicine
Pain Medicine   |   October 2009
Mortality Associated with Implantation and Management of Intrathecal Opioid Drug Infusion Systems to Treat Noncancer Pain
Anesthesiology 10 2009, Vol.111, 881-891. doi:10.1097/ALN.0b013e3181b64ab8
Anesthesiology 10 2009, Vol.111, 881-891. doi:10.1097/ALN.0b013e3181b64ab8
INITIATION and management of intrathecal opioid therapy involve the transition from systemic to intrathecal drug dosing, and sometimes to intrathecal plus systemic dosing. Long-term intrathecal infusion of the approved formulation of preservative-free morphine sulfate (Infumorph® and Duramorph®; Baxter Healthcare, Deerfield, IL) by using implantable pump and catheter systems is a medically accepted, US Food and Drug Administration (FDA)–approved, and Medicare- or private insurance-reimbursed therapy for patients with otherwise chronic intractable cancer or noncancer pain. The current work focuses on intrathecal opioid therapy for chronic noncancer pain. Authoritative data regarding intrathecal opioid versus  nonopioid intrathecal drug use are not available; however, information about intrathecal drug use in insurance databases and publications, including surveys of practitioners, reveals that exclusive use of nonopioid intrathecal analgesic drugs (e.g.  , without an opioid drug admixture) is infrequent in the Unites States.1–7 The intrathecal morphine sulfate label (package insert) contains drug dosage and patient care recommendations to help manage this therapy safely. The results of a randomized controlled clinical trial that was monitored for drug-related toxicity as a safety endpoint support the use of intrathecal opioid therapy for cancer pain.8 However, patients with noncancer pain presently comprise the majority of individuals implanted with intrathecal drug administration systems. Although the medical literature contains case series, review articles, consensus statements, and practice guidelines pertaining to this population, systematic safety, and efficacy data that meet a level of evidence comparable to that for cancer pain do not exist.1–6 
Voluntary postmarket reports to Medtronic (Minneapolis, MN) of three patient deaths within 1 day after implant in February 2006 was the initial signal that postimplant mortality among intrathecal opioid patients with noncancer pain might be higher than previously appreciated. Regulatory affairs, postmarket vigilance, and medical advisory personnel formed an ad hoc  team that identified a time-based cluster of early deaths by using a search strategy described in the Materials and Methods section. Medical review of the available data – including telephone interviews with implanting physicians and medical examiners–suggested that those deaths were attributable to overdoses of intrathecal opioids and/or systemic drugs. A preliminary statistical analysis found that early postimplant deaths were more frequent in noncancer pain patients treated with intrathecal opioids than in the first control samples that we identified–in particular, compared to patients treated with spinal cord stimulation. Only a small part of the mortality difference could be explained by age and sex differences. Early in the investigation, we transmitted mortality case reports to the FDA as MedWatch reports. We also initiated contact with the agency to share the results of our initial analyses before formally communicating information to physicians in a Dear Doctor letter in 2006. One of the authors (Dr. Owens) led the epidemiologic analyses described in this article. Thereafter, we obtained deidentified data from United Healthcare and Medicare, and we analyzed data samples matched for age and sex to explore potential differences between intrathecal opioid and spinal cord stimulation therapy patients that might explain the differences in mortality. Those analyses could not explain increased mortality in the intrathecal opioid therapy population, especially within three days of implant. We then examined other procedures (intrathecal drug dose changes, pump programming and refills, and catheter interventions) that do not apply to spinal cord stimulation or other pain therapies – but that turned out to be temporally associated with mortality in intrathecal opioid therapy patients. Our analyses, as outlined here and detailed in the Materials and Methods, Results, and Discussion sections, suggest that a number of clinical factors likely caused or contributed to patient mortality; despite limitations in the data and our epidemiological approach, communication of these findings to physicians may help to reduce patient mortality.
Materials and Methods
We used an epidemiological approach to investigate mortality rates after intrathecal opioid drug delivery system implants in noncancer pain patients by using complementary data sources and statistical methods described herein. The principal objective that evolved during this project was to compare the observed rates of mortality at different intervals after intrathecal opioid drug delivery system implant or other delivery system-related events by using mortality rates in other populations as benchmarks–including mortality in patients who underwent open spine surgery or implantation of spinal cord stimulation systems. Depending on the particular source, different categories of data were collected and analyzed as summarized in table 1, which also reflects the temporal sequence of steps in the investigation.
Table 1. Information and Analyses for Each Data Source 
Image Not Available
Table 1. Information and Analyses for Each Data Source 
×
No single database contained all of the information required to complete this project. We first focused on two internal data sources, the Enterprise Product Comment Report system (ePCR) and the Device Registration System (DRS), which are discussed in detail in the Index Cases and Product Comment Report System, and in the Device Registration System and Social Security Death Master File subsections. DRS data are matched on a monthly basis with the Social Security Administration (SS) Death Master File to provide a snapshot of patient survival after implantation of Medtronic devices. SS data are available to the medical device industry for purchase by subscription. However, ePCR, DRS, and SS data alone did not provide sufficient demographic or medical history information to compare mortality among matched patients treated by using different medical, surgical, or implant-based pain therapies–or versus  other control populations. Consequently, we purchased more detailed data from the Centers for Medicare and Medicaid Services (CMS) and from United Healthcare (UHC). Those data permitted us to analyze control populations on the basis of demographics, medical comorbidities, drug prescriptions, and healthcare utilization.9–11 Those data also permitted matched analyses of risk factors and mortality in cohorts with chronic pain treated medically, with an implanted intrathecal opioid drug delivery system, or with spinal cord stimulation–another implantable therapy for chronic pain. Summary results of the CMS and UHC data are discussed in this article, and detailed methods and results are presented in Supplemental Digital Content 1 (see document, which describes detailed methods and results, ), tables 1–4. We reviewed the published results of clinical trials, case series, consensus statements and surveys of intrathecal pain therapies,12–21 spinal cord stimulation,22–45 and population-based reports of mortality after conventional spine surgery.46–49 The aim of those ancillary analyses was to clarify whether the mortality rates observed at different intervals after initiation of intrathecal opioid therapy were related to the therapy itself or the result of confounding factors such as age, gender, underlying medical conditions, or prescription drugs.
Table 2. Summary of Nine Index Cases of Death within 3 Days after Intrathecal Opioid Device Implant 
Image Not Available
Table 2. Summary of Nine Index Cases of Death within 3 Days after Intrathecal Opioid Device Implant 
×
Table 2. Continued 
Image Not Available
Table 2. Continued 
×
Table 3. Unadjusted Death Rates for Benchmark Therapies: Lumbar Diskectomy, Lumbosacral Spine Surgery, Spinal Cord Stimulation Implantation, and Intrathecal Drug Delivery Implantation for Noncancer Pain 
Image Not Available
Table 3. Unadjusted Death Rates for Benchmark Therapies: Lumbar Diskectomy, Lumbosacral Spine Surgery, Spinal Cord Stimulation Implantation, and Intrathecal Drug Delivery Implantation for Noncancer Pain 
×
Table 4. Deaths/Expected Deaths Ratios for Intrathecal Opioid and Spinal Cord Stimulation Patients by Time Period 
Image Not Available
Table 4. Deaths/Expected Deaths Ratios for Intrathecal Opioid and Spinal Cord Stimulation Patients by Time Period 
×
Index Cases and Product Comment Report System
The investigation began in the first 10 days of February 2006 when we received three voluntary reports from health professionals of mortality within 24 h after implantation of an intrathecal drug delivery system to treat noncancer pain. These are case numbers 1, 6, and 8 in table 2and were recognized as a signal that further investigation was warranted. The next step was a prospectively defined and time-bounded search of the internal postmarket safety database (ePCR) for additional early postimplant death cases. The time boundaries were from December 1, 2005 (2 months before the first reports) to March 31, 2006 (almost 2 months after the first reports).
The ePCR system collects all spontaneous adverse event reports for Medtronic products, whether drug-related or device-related, from all sources. In a sense, ePCR is a complaint file, and all cases that involve mortality or other serious consequences as defined by FDA are reported through the MedWatch system. Categories of information collected through ePCR are summarized in table 1. Whenever possible, identifying patient data are collected in order to integrate data with the Device Registration System (described in the Device Registration System and Social Security Death Master File subsection). Limitations of voluntary reporting in ePCR include incomplete information captured for different cases, for example, on the cause of death, concomitant medications, or other medical data. ePCR reports are voluntary; therefore, the number of events in this database is an underestimate of what actually occurs. The reason to search for cases in the 4 months surrounding the three sentinel cases was to increase the likelihood of detecting a relationship between early mortality and potential drug or device manufacturing issues based on dates or lot numbers. We prespecified a 3-day postimplant window as the working definition for early mortality to account for delayed discovery or uncertainties regarding the time of death for unwitnessed cases. Table 2summarizes the findings in nine cases between December 2005 and March 2006. These include the three sentinel cases (numbers 1, 6, and 8) plus six more cases (numbers 2, 3, 4, 5, 7, and 9) identified by the search strategy. Active investigation of these cases revealed more individual patient-, medical care-, drug-, and device-related data, and postmortem findings than are ordinarily captured through spontaneous, voluntary reporting activities to FDA.
Device Registration System and Social Security Death Master File
The Medtronic DRS also collects data mandated by FDA and the Code of Federal Regulations for implantable medical devices. Information captured by DRS is summarized in table 1, and it is patient-specific, identifiable, and in contrast to ePCR, prospectively collected. DRS data are available for greater than 90% of patients implanted in the United States to treat chronic noncancer pain with Medtronic intrathecal opioid delivery or spinal cord stimulation systems. Matching of DRS data and the SS Death Master File—a database that gathers information reported to state vital records offices—allowed us to compare mortality rates at selected intervals after implantation with either an intrathecal opioid drug delivery or spinal cord stimulation device. In 2006, we prospectively established the time window to investigate postimplant mortality with either device as any implant that occurred between May 1998 and December 2004 to allow 1-yr follow-up through January 1, 2006. Time intervals chosen for analysis were in the first 3 days, in the first 30 days, and in the first year after implant. Age and gender-adjusted comparisons of death rates for patients with intrathecal opioid delivery and spinal cord stimulation systems employed direct standardization to the combined populations of all patients with either device. We assumed for the purpose of this analysis that devices remained in use for at least 1 yr (no censoring) because of the limitation that DRS does not always record whether or when devices in living patients are explanted or allowed to remain unused. An additional limitation is that DRS records include the indication for intrathecal opioid therapy (e.g.  , noncancer pain) but not the intrathecal or systemic medications being used. However, data about intrathecal drugs and concomitant medications were available for analyses of the other data sources listed in table 1and discussed in the Results section.
Patients implanted with spinal cord stimulation devices served as one of the benchmark control groups for this and other analyses because, despite certain differences, patients treated by using intrathecal opioid or spinal cord stimulation therapies most commonly suffer from chronic noncancer pain that does not respond adequately to other therapies, including systemic opioid drugs. The surgical implant procedures for drug delivery and spinal cord stimulation are similar to each other, and neither involves surgical entry into a body cavity. Moreover, mortality data for the two devices in DRS are expected to be equally complete. For the cohorts with each device, the expected numbers of deaths over the 3-day, 30-day, or 1-yr time intervals were calculated on the basis of published actuarial tables for the overall US population by age and gender. The ratio of observed to expected deaths is the standardized mortality ratio, and confidence limits for these ratios were calculated by using Byar's method.50 The age and gender distributions of the populations with intrathecal opioid delivery and spinal cord stimulation systems are different; for comparison of the mortality experience of these two populations, the method of direct standardization was used, with weights calculated from the overall distribution of subjects with either device.
Additional Analyses–Product Comment Report System
Examination of the nine index cases and the comparison of mortality rates between intrathecal opioid patients and spinal cord stimulation patients suggested possible explanations for 3-day postimplant mortality. Preliminary analyses of 30-day and 1-yr mortality also suggested a longer-term excess in mortality among intrathecal opioid patients compared to spinal cord stimulation and other populations. Consequently, we performed additional analyses on 1,851 ePCR reports for intrathecal opioid therapy implants and mortality that occurred during the same interval as the DRS analysis (described in the preceding two paragraphs). The additional analyses of ePCR complaint files evaluated deaths reported to have occurred within 3 days of a new intrathecal opioid therapy system implant and expanded to include deaths that occurred within 3 days after any pump refill, pump replacement, catheter revision, pump programming session, or drug dose change. A limitation of this additional analysis is that the date of death–with respect to device implants, replacements, revisions, refills, or programming–was reported for only 36% of noncancer pain cases (201 of 557). Within limitations imposed by the reported data, the analysis revealed temporal relationships between patient deaths and the listed interventions.
Results
Index Cases
The clinical features of the nine cases identified between December 2005 and March 2006 and the medical circumstances surrounding their deaths are summarized in table 2. Analysis of clinical, postmortem, and toxicological data indicate that the cause of death involved opioid and/or central nervous system depressant drug overdose as a primary or contributing factor in every case. Eight of the nine patients had noncancer pain, and one patient had cancer pain. Eight of the nine patients, seven with noncancer pain plus the one with cancer pain, died within 24 h after a new device implant, pump replacement, or catheter revision. One of the nine patients died within 48 h after hospital discharge. Seven cases underwent new intrathecal opioid delivery system implants, and two had intrathecal opioid delivery restored weeks or longer after catheter dislodgement or temporary pump removal–the latter performed for a mistaken diagnosis of pump pocket infection. Seven cases (numbers 1, 2, 3, 5, 6, 7, and 9) were programmed to receive an intrathecal morphine or morphine-equivalent dosage of hydromorphone that exceeded the starting dose for intrathecal opioid naïve patients recommended in the Infumorph labeling (recommended starting dose = 0.2–1.0 mg/d for individuals without opioid tolerance). The other two cases (numbers 4 and 8) were programmed to receive less than 1.0 mg/d intrathecal morphine (0.99 and 0.75 mg/d, respectively), but they died from respiratory arrest resulting from apparent intrathecal opioid or mixed intrathecal opioid plus concomitant systemic drug overdose after hospital or surgical center discharge. Errors in dosage calculations or pump programming caused or contributed to two deaths (cases 7 and 9). Case 9 died within hours after a new system implant and initiation of therapy with hydromorphone and baclofen.
Apart from the medical causes and circumstances of death for these nine cases, we also performed analyses to determine whether this cluster of events was associated with time-, drug-, or device-related factors. Data were segmented according to the date of surgery or death, by date of device manufacture, and by device serial or lot number. None of the device analyses revealed a clear signal or trend, and the intrathecal drugs administered were from multiple manufacturers or pharmacies. Thus, the nine cases from December 2005 to March 2006 did not constitute an outbreak or epidemic that could be traced to a particular source or cause. Rather, the cases constituted a coincidental temporal cluster within the large data set of intrathecal opioid therapy implants for noncancer pain.
Device Registration System and Social Security Death Master File
The mortality rates calculated from DRS and SS data within 3 days, 30 days, and 1 yr after intrathecal opioid delivery system implant are summarized in table 3. Corresponding DRS/SS data for mortality after implant of a spinal cord stimulation system are also shown. For benchmarking purposes, published in-hospital mortality rates after lumbar diskectomy in a nationwide community hospital sample43,44 and mortality after a variety of lumbosacral spine operations in Medicare beneficiaries45 are also provided. In-hospital mortality after lumbar spine surgery varied with the population under study and the complexity of the operations performed. The lowest mortality rate, 0.059% (0.59 per 1,000 procedures) occurred among diskectomy patients in the community hospital setting. Hospital mortality among older Medicare beneficiaries who underwent more complex operations, such as multilevel surgery and/or fusion, was an order of magnitude higher: 0.52% (5.2 per 1,000 procedures). Thirty-day and 1-yr mortality in the Medicare population also are presented in table 3.
Mortality within 3 days after intrathecal opioid system implantation was 0.088% (0.88 per 1,000)–higher than the community hospital diskectomy population (0.59 per 1,000) but lower than the Medicare spine surgery population (5.2 per 1,000). The 0.088% 3-day intrathecal opioid mortality rate also was eight times greater than the 0.011% (0.11 per 1,000) 3-day mortality rate after spinal cord stimulator implantation. The mortality rate in the intrathecal opioid population remained higher, albeit by lower proportions, than in the spinal cord stimulation population at 30 days and 1 yr after implant. Standardization to the overall age and sex distribution of the population implanted with either system makes the intrathecal opioid therapy and spinal cord stimulation rates directly comparable without confounding by age and sex differences (e.g.  , average age higher by 4.6 yr at implant for intrathecal opioid compared to spinal cord stimulation patients). This suggests that the observed excess mortality was attributable to intrathecal opioid therapy and not to age or gender differences. The fourth row of table 3shows the standardized mortality ratios for intrathecal opioid/spinal cord stimulation patients. Those ratios are only slightly smaller than the unadjusted (for age and sex) ratios of rates in the third row of table 3, again indicating that age and gender differences accounted for only a small portion of the difference in mortality observed in this population for the two therapies. Confidence intervals for these ratios exclude unity, indicating that the excess mortality among intrathecal opioid patients is highly statistically significant.
Table 4shows the ratios of the number of deaths in the intrathecal-opioid cohort and spinal cord stimulation cohort to the number of deaths that would have been expected on the basis of the overall age- and gender-specific mortality rates for the US population in 2002 (a ratio of 1.0 would indicate no excess mortality). The ratios of observed deaths for noncancer intrathecal opioid therapy patients to the number of expected deaths (7.5:1 at 3 days, 3.4:1 at 30 days, and 2.7:1 at 1 yr) remained higher at all intervals compared to spinal cord stimulation patients (1.4:1, 1.1:1, and 1.4:1, respectively). Confidence limits for the spinal cord stimulation ratios are close to or include unity, indicating that mortality was not greater than expected for spinal cord stimulation patients.
Product Comment Report Database and Returned Product Analysis
Evaluation of 557 ePCR reports of noncancer pain intrathecal opioid therapy patient deaths between 1998 and 2007 revealed that 88 cases (15.8%) were reported to have occurred within 3 days of a pump implant, refill, replacement, reprogramming, dose change, or catheter revision (table 5). Another 113 deaths occurred more than 3 days after any of those procedures for a total of 201 cases (36% of 557) with a known time of death. A reported cause of death was available for 55 (62.5%) of the 88 death cases within 3 days of an implant, refill, replacement, reprogramming, dose change, or catheter revision as summarized in table 6. Suspected or confirmed drug overdose was the most frequently reported cause reported by physicians (25/55 cases). When data are tabulated by the type of device-related procedure (table 7), deaths within 3 days of an initial implant or pump replacement account for half of the observed mortality, despite the infrequency of these procedures relative to the number of refills and dose changes. These data include spontaneously reported deaths, only—in contrast to the more comprehensive DRS-SS analysis described in the preceding paragraph. One should interpret the ePCR data cautiously because of that limitation.
Table 5. Time of Death in Noncancer Pain Patients in Relation to a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
Image Not Available
Table 5. Time of Death in Noncancer Pain Patients in Relation to a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
×
Table 6. Reported Cause of Death in Noncancer Pain Patients within 3 Days of a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
Image Not Available
Table 6. Reported Cause of Death in Noncancer Pain Patients within 3 Days of a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
×
Table 7. Noncancer Pain Patient Deaths within 3 Days by Pump- or Device-related Procedure–From Enterprise Product Comment Report System 
Image Not Available
Table 7. Noncancer Pain Patient Deaths within 3 Days by Pump- or Device-related Procedure–From Enterprise Product Comment Report System 
×
Pumps were returned for analysis in 41 ePCR mortality cases. Analysis revealed an abnormality in 8 (19.5%) of 41 cases. Deidentified raw data for those eight cases are provided in table 5, Supplemental Digital Content 1 (see document, which is a table summarizing device analysis results, ). Abnormal findings included battery depletion/end of life (n = 2), pump memory errors (n = 2), reservoir septum damage (n = 2), broken motor screws (n = 1), and bridging residue on gears (n = 1). Battery depletion/end of life means that the pump is stopped and no longer dispensing fluid. Pump memory error is a programmer display that happens when the most recent (or current) programming session is unsuccessful because of background interference, most commonly if the device and programmer are too close to a television or computer screen, or if programming is attempted postmortem or after explant when the pump (or body) is at room temperature. A memory error does not indicate pump damage, but the pump may stop or revert to the previously programmed settings. Reservoir septum damage is a visual observation of the refill port septum, which indicates only that the pump has been refilled many times, but has no influence on pump function. Broken motor screws and bridging residue on gears are caused by use of specific off-label drugs—for example, meperidine (United States), diacetyl morphine (United Kingdom)—that are chemically incompatible with pump components, and eventually leak into the pump mechanism.6 These ultimately cause the pump to stall, with the stalled condition displayed on the programmer screen during the next physician visit. However, the two affected pumps described here had not yet stalled. Abnormal observations in these eight devices could not directly cause patient mortality from intrathecal drug overdose because the findings either had no influence on flow rate accuracy, or caused the affected pumps to cease infusion rather than to overinfuse the intrathecal drugs.
The 33 returned pumps (80.5%) that showed no anomalies included all of the pumps in the ePCR database that were explanted postmortem from individuals who died within a few hours or days after implantation. Returned product analysis and/or interrogation and aspiration of pump contents by medical examiners in five of the nine index cases also revealed that the expected versus  actual residual volume of drug in the pump reservoir was within device accuracy specifications. In three of the index cases, pump performance data were unavailable because the device was not removed before burial, and in one additional case because the medical examiner would not disclose information.
Discussion
Limitations of the Analyses
The data, analyses, and conclusions of this investigation are subject to several limitations. As described in the Methods section, active investigation of the nine index cases provided detailed data that are not ordinarily captured by passive processes for spontaneously reported adverse events. Therefore, in contrast to the nine index cases, analyses conducted on the larger set of ePCR data reported by healthcare professionals or family members between 2005 and 2007 were intentionally limited to date of death, reported cause of death, and temporal relationship to a pump implant or replacement or to other procedure as described in the Results section and tables 5 and 6. Those analyses were further limited by gaps in the spontaneously reported data. A time of death was not reported in 64% of ePCR death cases (356 of 557), and a cause of death was not reported in 37.5% of mortality cases (33 of 88) that occurred within 3 days of a pump or system procedure.
Investigation of the nine index cases (table 2) also left gaps in the data, especially regarding the sources of intrathecal drugs and postmortem analysis of the pump reservoir contents. Among the six cases receiving intrathecal morphine (all as monotherapy), two were administered a parenteral morphine formulation not labeled for intrathecal use (Hospira Healthcare, Inc., Lake Forest, IL), two were administered morphine for which the drug source (commercial or compounded) could not be determined, and one case each was administered hospital-compounded morphine sulfate or the approved formulation of Infumorph. Among the three cases treated by using hydromorphone (monotherapy in one case, admixture in two cases), one received Dilaudid (Knoll Pharmaceutical Co., Mt. Olive, NJ), and the drug source could not be determined in two cases. In three cases, the concentration of hydromorphone (case 1, unknown source) or morphine (cases 3 and 5, Hospira) were measured postmortem and found to be lower than what the implanting physicians had programmed into the pump. It is possible, albeit speculative, that pharmacists made drug compounding or dilution errors in these individual cases. Another explanation for lower than expected reservoir concentrations is that implanters were unable to completely aspirate all of the sterile water that remains in the pump reservoir after manufacturing. This manufacturing step ensures that pumps can be factory-set to run at a minimal shelf state and not stall during prolonged shelf life. It is also possible that both mechanisms contributed to the lower observed drug concentrations in these cases. Table 2notes that two of the index cases (cases 4 and 8) were obese. We could not examine whether body habitus or obesity contributed to postanesthetic or delayed patient deaths in other populations because body weight or body mass index values were not included in the data sets.
Device stoppage or underinfusion was unlikely to have been followed by uncontrolled pain and fatally excessive oral medication intake among the nine index cases or in seven of eight of the 3-day postimplant cases identified in the 2005–2007 ePCR analysis. The patients in both data sets died before experiencing intrathecal drug-mediated analgesic effects for a long enough period to have experienced pain recurrence before death. However, it is theoretically possible that case 3 (battery end of life) in table 5, Supplemental Digital Content 1 (see document, a table summarizing device analysis results, ) or other patients who died after months or years on intrathecal opioid therapy—and who had stopped or underinfusing pumps identified at the time of death—could have unintentionally overdosed themselves on systemic medications because of recurrent pain. Among the 19.5% of pumps (8 of 41 returns in 557 deaths in ePCR) that had anomalies detected by returned product analysis, two pumps had dead batteries and definitely were nonfunctioning, and another two were in a “pump memory error” state that can cause pump stoppage with clinical consequences if stoppage occurs before death. However, a pump memory error also can occur if a pump is interrogated during postmortem investigations or after being returned to a physician after explant by mortuary personnel. The available data do not reveal when those pumps entered the memory error state—in life or after explant (table 5, Supplemental Digital Content 1, which summarizes device analysis results, ).
The Device Registration System captures individual patient identification data, including age and sex, implanting physician and insitution data, the serial and/or lot number(s) of the implanted device(s), the date of implant, and the patient's underlying diagnosis or indication for device implantation. DRS does not collect any other medical or health-related information. The SS Death Master File provides the identity and date of death of deceased individuals, but it does not report the cause of death or other medical information. The DRS and SS data correlation that we performed for patients with noncancer pain implanted with intrathecal opioid delivery or spinal cord stimulation systems demonstrated a clear excess in early deaths among intrathecal opioid patients that could not be explained by differences in age and gender, but the data could not rule out other potential differences between these cohorts that might be explanatory. Our DRS data do not include any information on medical history or other factors that could potentially explain the differences in mortality for these two cohorts; we therefore purchased data from the CMS and UHC to specifically analyze whether differences in underlying medical comorbidities, physician-prescribed drugs, healthcare utilization, as well as age and sex, could explain any of the observed mortality difference between noncancer pain patients treated by using intrathecal opioid and spinal cord stimulation therapy (methods described in detail in Supplemental Digital Content 1 [see document, ]). Therapy-specific data from CMS and UHC (intrathecal opioids, spinal cord stimulation, or no invasive therapy) permitted similar analyses as the DRS-SS data but included devices made by all manufacturers and with the ability to control for demographic and medical comorbidities.
Mortality was the endpoint that we investigated on the basis of an initial safety signal and a temporal cluster of events. Nonfatal procedure- or therapy-related injuries that lead to disability or potential loss of life years afterwards are other important topics that deserve examination in the future. A discussion of other limitations and mitigating factors pertaining to analyses of the CMS and UHC data sets also appears in Supplemental Digital Content 1 (see document, ).
Interpretation and Implications of the Results
Analyses of data from multiple complementary internal and population-based sources indicate that noncancer pain patients treated with intrathecal opioid therapy experience elevated 3-day, 30-day, and 1-yr mortality compared to patients treated by using spinal cord stimulation, another device-based therapy that involves a similar, albeit not identical, implant procedure. Intrathecal opioid therapy patients also experience slightly higher 3-day mortality compared to in-hospital deaths for community hospital lumbar diskectomy patients, and intrathecal opioid patients experience higher 30-day and 1-yr mortality (but lower 3-day or in-hospital mortality) than Medicare patients who undergo more complex spinal operations. Although a larger proportion of intrathecal opioid implant procedures compared to spinal cord stimulation procedures are performed under general anesthesia, none of the demographic, concomitant illness, or other healthcare utilization data that we reviewed suggest that increased anesthetic risks were responsible for the mortality differences that we observed. Absolute and comparative mortality rates for intrathecal opioid therapy initiation and maintenance had not been identified or explored systematically before this investigation. One possible reason for underappreciation of the risks that we identified include the low likelihood that an individual physician would observe early postimplant mortality, which occurs at a rate of fewer than one per 1,000 device implant or component replacement procedures.
This investigation began when three sentinel reports in February 2006 and six subsequently identified nearly contemporaneous cases (nine index cases) suggested that intrathecal opioids, or mixed intrathecal opioid plus systemic drug interactions, caused fatal respiratory depression. Available data from the index cases, and from a follow-on investigation in the ePCR data base, albeit subject to limitations described in the opening paragraphs of this Discussion section, revealed that in cases where pump volumes or performance were checked postmortem, device-related malfunctions or overinfusion did not cause intrathecal overdosage and death. One important limitation of this finding that bears repeating is that fewer than 10% of death cases in the ePCR database were accompanied by returned products for analysis. Therefore, one cannot infer that device malfunctions or overinfusion events were absent in the 90% of deaths where residual volumes were not analyzed or devices were not returned. The major finding from the index cases is that early deaths do occur as an apparent result of intrathecal and systemic opioid overdose and that all spontaneously-reported early deaths in the 4-month analysis window were variations on this theme.
The DRS-SS analysis revealed that the excess mortality risk of intrathecal opioid mortality versus  spinal cord stimulation was 7.56:1 at 3 days, 3.64:1 at 30 days, and 2.25:1 at 1-yr postimplant, after adjustment for age and gender. Although the relative risk is greatest shortly after implant, the largest number of deaths occurs during therapy maintenance. Analysis of spontaneously reported ePCR data beyond the index cases, despite the limitations that we have enumerated, suggested a temporal relationship between mortality and pump refill or programming procedures (in addition to implant or replacement procedures). That analysis suggests, but does not prove, that elevated 30-day and 1-yr mortality among intrathecal opioid therapy patients may also be therapy-related.
Large population-based analyses point in the same direction as the internal data; intrathecal opioid therapy patients experience excess mortality risks versus  control- or matched control populations treated differently. Those same sources, especially the UHC cohorts, also did not support the anecdotal notion that intrathecal opioid patients, when matched for age and gender, are sicker, are at higher risk for nontherapy-related death, or have more drugs prescribed than spinal cord stimulation or nonimplanted patients. Intrathecal opioid patients also must undergo pump refills, each of which carries additional risks, as identified by the persistently elevated 30-day and 1-yr mortality compared to spinal cord stimulation patients, who only experience additional device risks if they undergo a revision or replacement procedure.
The finding that most of the mortality occurred out of hospital in each data set that we analyzed dovetails with recent publications and news reports that described unexpectedly high outpatient mortality rates associated with systemic opioid prescriptions accompanied by other medication and respiratory depressant substance intake, including alcohol and prescribed or illicit drugs.51–53 Removing the stigma for pain patients and lowering the barriers for physicians to prescribe chronic opioid analgesics undoubtedly has reduced pain and suffering. Those benefits are accompanied by risks. In a recent population-based study of mortality after lumbar spine fusion surgery, analgesic-related deaths were the single largest cause of mortality, accounting for 21% of deaths and 31.4% of potential life lost.46 Still, one cannot infer that because nine index cases died from apparent intrathecal opioid or intrathecal opioid plus systemic drug-induced respiratory depression that the other intrathecal therapy death cases in databases that we examined (DRS, ePCR, CMS, or UHC) also must have died from drug-induced respiratory depression. Rather, all of the analyses identified excess mortality among intrathecal opioid patients, that excess mortality was at least partly therapy-related, and that a candidate for further investigation is intrathecal or intrathecal plus systemic drug-induced respiratory depression. Even a cautious interpretation of these results raises the question of what physicians and industry can do now, albeit based on incomplete and imperfect information, to reduce the preventable proportion of mortality associated with intrathecal opioid therapy for noncancer pain. Physicians may wish to critically examine the prescribing information, precautions (including special populations), and warnings contained in the Infumorph and Duramorph labeling. Concerns about potentially fatal respiratory depression after even small doses of intrathecal morphine have a sound physiologic and experimentally demonstrable basis.54,55 
Morphine sulfate labeling and safety information should apply to hydromorphone (from 3:1 to 6:1 higher potency than morphine, depending on the reference56) and other opioids, even though the latter are not FDA-approved for chronic intrathecal administration. With respect to device-related aspects of therapy, two patient deaths among the nine index cases were caused by pump programming and/or intrathecal dose calculation errors. This suggests that improvements in physician education and in the device interface–including greater simplicity, and improved device-generated alerts–also are important changes to enhance patient safety.
Conclusions
Despite limitations of the data and analyses for each component of this investigation, the finding of excess early and delayed mortality associated with intrathecal opioid therapy for noncancer pain appears genuine. Physicians and industry are now confronted with the challenge and opportunity to reduce the preventable portion of the excess mortality described in this article. Further investigations, educational efforts, and–ultimately–changes in medical practices may be required.
References
Bennett G, Serafini M, Burchiel K, Buchser E, Classen A, Deer T, Du Pen S, Ferrante FM, Hassenbusch SJ, Lou L, Maeyaert J, Penn R, Portenoy RK, Rauck R, Willis KD, Yaksh T: Evidence-based review of the literature on intrathecal delivery of pain medication. J Pain Symptom Manage 2000; 20:S12–36Bennett, G Serafini, M Burchiel, K Buchser, E Classen, A Deer, T Du Pen, S Ferrante, FM Hassenbusch, SJ Lou, L Maeyaert, J Penn, R Portenoy, RK Rauck, R Willis, KD Yaksh, T
Bennett G, Deer T, Du Pen S, Rauck R, Yaksh TL, Hassenbusch SJ: Future directions in the management of pain by intraspinal drug delivery. J Pain Symptom Manage 2000; 20:S44–50Bennett, G Deer, T Du Pen, S Rauck, R Yaksh, TL Hassenbusch, SJ
Deer TR, Krames E, Levy RM, Hassenbusch SJ, Prager JP: Practice choices and challenges in the current intrathecal therapy environment: An online survey. Pain Med 2009; 10:304–9Deer, TR Krames, E Levy, RM Hassenbusch, SJ Prager, JP
Deer T, Krames E, Hassenbusch S, Prager JP: Polyanalgesic consensus conference 2007: Recommendations for the management of pain by intrathecal (intraspinal) drug delivery: Report of an interdisciplinary expert panel. Neuromodulation 2007; 10:300–28Deer, T Krames, E Hassenbusch, S Prager, JP
Ellis DJ, Dissanayake S, McGuire D, Charapata SG, Staats PS, Wallace MS, Grove GW, Vercruysse P, the Elan Study 95-002 Group: Continuous intrathecal infusion of ziconotide for treatment of chronic malignant and nonmalignant pain over 12 months: A prospective, open-label study. Neuromodulation 2008; 11:40–9Ellis, DJ Dissanayake, S McGuire, D Charapata, SG Staats, PS Wallace, MS Grove, GW Vercruysse, P the Elan Study 95-002 Group,
Hassenbusch SJ, Portenoy RK, Cousins M, Buchser E, Deer TR, Du Pen SL, Eisenach J, Follett KA, Hildebrand KR, Krames ES, Levy RM, Palmer PP, Rathmell JP, Rauck RL, Staats PS, Stearns L, Willis KD: Polyanalgesic consensus conference 2003: An update on the management of pain by intraspinal drug delivery–report of an expert panel. J Pain Symptom Manage 2004; 27:540–63Hassenbusch, SJ Portenoy, RK Cousins, M Buchser, E Deer, TR Du Pen, SL Eisenach, J Follett, KA Hildebrand, KR Krames, ES Levy, RM Palmer, PP Rathmell, JP Rauck, RL Staats, PS Stearns, L Willis, KD
Rauck R, Wallace M, Leong M, Minehart M, Webster LR, Charapata SG, Abraham JE, Buffington DE, Ellis D, Kartzinel R: Ziconotide 301 Study Group: A randomized double-blind, placebo-controlled study of intrathecal ziconotide in adults with severe chronic pain. J Pain Symptom Manage 2006; 31:393–406Rauck, R Wallace, M Leong, M Minehart, M Webster, LR Charapata, SG Abraham, JE Buffington, DE Ellis, D Kartzinel, R
Smith TJ, Staats PS, Deer T, Stearns LJ, Rauck RL, Boortz-Marx RL, Buchser E, Catala E, Bryce DA, Coyne PJ, Pool GE: Randomized clinical trial of an implantable drug delivery system compared with comprehensive medical management for refractory cancer pain: Impact on pain, drug-related toxicity, and survival. J Clin Oncol 2002; 20:4040–9Smith, TJ Staats, PS Deer, T Stearns, LJ Rauck, RL Boortz-Marx, RL Buchser, E Catala, E Bryce, DA Coyne, PJ Pool, GE
Charlson ME, Pompei P, Ales KL, McKenzie CR: A new method of classifying prognostic comorbidity in longitudinal studies: Development and validation. J Chron Dis 1987; 40:373–83Charlson, ME Pompei, P Ales, KL McKenzie, CR
Klabunde C, Potosky A, Legler J, Warren J: Development of a comorbidity index using physician claims data. J Clin Epidem 2000; 53:1258–67Klabunde, C Potosky, A Legler, J Warren, J
Klabunde C, Warren J, Legler J: Assessing comorbidity using claims data – an overview. Med Care 2002; 40(8 Suppl):26–35Klabunde, C Warren, J Legler, J
Anderson V, Burchiel K: A Prospective study of long-term intrathecal morphine in the management of chronic nonmalignant pain. Neurosurgery 1999; 44:289–300Anderson, V Burchiel, K
Brown J, Klapow J, Doleys D, Lowery D, Tutak U: Disease-specific and generic health outcomes: A model for the evaluation of long-term intrathecal opioid therapy in noncancer low back pain patients. Clin J Pain 1999; 15:122–31Brown, J Klapow, J Doleys, D Lowery, D Tutak, U
Deer T, Chapple I, Classen A, Javery K, Stoker V, Tonder L, Burchiel K: Intrathecal drug delivery for treatment of chronic low back pain: Report from the national outcomes registry for low back pain. Pain Med 2004; 5:6–13Deer, T Chapple, I Classen, A Javery, K Stoker, V Tonder, L Burchiel, K
Doleys DM, Brown J, Ness T: Multidimensional outcomes analysis of intrathecal, oral opioid, and behavioral–functional restoration therapy for failed back surgery syndrome: A retrospective study with 4 years' follow-up. Neuromodulation 2006; 9:270–83Doleys, DM Brown, J Ness, T
Follett KA, Boortz-Marx RL, Drake JM, DuPen S, Schneider SJ, Turner MS, Coffey RJ: Prevention and management of intrathecal drug delivery and spinal cord stimulation system infections. Anesthesiology 2004; 100:1582–94Follett, KA Boortz-Marx, RL Drake, JM DuPen, S Schneider, SJ Turner, MS Coffey, RJ
Rainov N, Heidecke V, Burkurt W: Long term intrathecal infusion of drug combinations for chronic back and leg pain. J Pain Symptom Manage 2001; 22:862–71Rainov, N Heidecke, V Burkurt, W
Shaladi A, Saltari MR, Piva B, Crestani F, Tartari S, Pinato P, Micheletto G, Dall'Ara R: Continuous intrathecal morphine infusion in patients with vertebral fractures due to osteoporosis. Clin J Pain 2007; 23:511–7Shaladi, A Saltari, MR Piva, B Crestani, F Tartari, S Pinato, P Micheletto, G Dall'Ara, R
Staats P, Whiteworth M, Barakat M, Anderson W, Lilienfeld S: The use of implanted programmable infusion pumps in the management of non-malignant chronic low back pain. Neuromodulation 2007; 10:376–80Staats, P Whiteworth, M Barakat, M Anderson, W Lilienfeld, S
Thimineur MA, Kravitz E, Vodapally MS: Intrathecal opioid treatment for chronic non-malignant pain: A 3-year prospective study. Pain 2004; 109:242–9Thimineur, MA Kravitz, E Vodapally, MS
Winkelmuller M, Winkelmuller W: Long-term effects of continuous intrathecal opioid treatment in chronic pain of nonmalignant etiology. J Neurosurg 1996; 85:458–67Winkelmuller, M Winkelmuller, W
Allegri M, Arachi G, Barbieri M, Paulin L, Bettaglio R, Bonetti G, Demartini L, Violini A, Braschi A, Bonezzi C: Prospective study of the success and efficacy of spinal cord stimulation. Minerva Anestesiol 2004; 70:117–24Allegri, M Arachi, G Barbieri, M Paulin, L Bettaglio, R Bonetti, G Demartini, L Violini, A Braschi, A Bonezzi, C
Barolat G: A prospective multicenter study to assess the efficacy of spinal cord stimulation utilizing a multi-channel radio-frequency system for the treatment of intractable low back and lower extremity pain. Initial considerations and methodology. Neuromodulation 1999; 2:179–83Barolat, G
Bennett DS, Alo´ KM, Oakley J, Feler CA: Spinal cord stimulation for complex regional pain syndrome I [RSD]: A retrospective multicenter experience from 1995 to 1998 of 101 Patients. Neuromodulation 1999; 2:202–10Bennett, DS Alo´, KM Oakley, J Feler, CA
Burchiel KJ, Anderson VC, Brown FD, Fessler RG, Friedman WA, Pelofsky S, Weiner RL, Oakley J, Shatin D: Prospective, multicenter study of spinal cord stimulation for relief of chronic back and extremity pain. Spine 1996; 21:2786–94Burchiel, KJ Anderson, VC Brown, FD Fessler, RG Friedman, WA Pelofsky, S Weiner, RL Oakley, J Shatin, D
Dario A, Fortini G, Bertolo D, Bacuzzi A, Grizzetti C, Cuffari S: Treatment of failed back surgery syndrome. Neuromodulation 2001; 4:105–10Dario, A Fortini, G Bertolo, D Bacuzzi, A Grizzetti, C Cuffari, S
Devulder J, De Laat M, Bastelaere M, Rolly G: Spinal cord stimulation: A valuable treatment for chronic failed back surgery patients. J Pain Sympt Manage 1997; 13:296–301Devulder, J De Laat, M Bastelaere, M Rolly, G
Forouzanfar T, Kemler MA, Weber WE, Kessels AGH, van Kleef M: Spinal cord stimulation in complex regional pain syndrome: Cervical and lumbar devices are comparably effective. Br J Anaesth 2004; 9:348–53Forouzanfar, T Kemler, MA Weber, WE Kessels, AGH van Kleef, M
Harke H, Gretenkort P, Ladleif HU, Koester P, Rahman S: Spinal cord stimulation in postherpetic neuralgia and in acute herpes zoster pain. Anesth Analg 2002; 94:694–700Harke, H Gretenkort, P Ladleif, HU Koester, P Rahman, S
Kavar B, Rosenfeld JV, Hutchinson A: The efficacy of spinal cord stimulation for chronic pain. J Clin Neurosci 2000; 7:409–13Kavar, B Rosenfeld, JV Hutchinson, A
Kay AD, McIntyre MD, Macrae WA, Varma TRK: Spinal cord stimulation–a long-term evaluation in patients with chronic pain. Br J Neurosurg 2001; 15:335–41Kay, AD McIntyre, MD Macrae, WA Varma, TRK
Kim SH, Tasker RR, Oh MY: Spinal cord stimulation for nonspecific limb pain versus  neuropathic pain and spontaneous versus  evoked pain. Neurosurgery 2001; 48:1056–64Kim, SH Tasker, RR Oh, MY
Kemler MA, Barendse GA, van Kleef M, de Vet HC, Rijks CP, Furnee CA, van den Wildenberg FAJM: Spinal cord stimulation in patients with chronic reflex sympathetic dystrophy. N Engl J Med 2000; 343:618–24Kemler, MA Barendse, GA van Kleef, M de Vet, HC Rijks, CP Furnee, CA van den Wildenberg, FAJM
Kumar K, Taylor RS, Jacques L, Eldabe S, Meglio M, Molet J, Thomson S, O'Callaghan J, Eisenberg E, Milbouw G, Buchser E, Fortini G, Richardson J, North RB: Spinal cord stimulation versus  conventional medical management for neuropathic pain: A multicentre randomised controlled trial in patients with failed back surgery syndrome. Pain 2007; 132:179–88Kumar, K Taylor, RS Jacques, L Eldabe, S Meglio, M Molet, J Thomson, S O'Callaghan, J Eisenberg, E Milbouw, G Buchser, E Fortini, G Richardson, J North, RB
Lanner G, Spendel MC: Spinal cord stimulation for the treatment of chronic non-malignant pain. Acta Neurochir 2007; 97:79–84Lanner, G Spendel, MC
North RB, Ewend MG, Lawton MT, Kidd DH, Piantadosi S: Failed back surgery syndrome: 5-year follow-up after spinal cord stimulator implantation. Neurosurgery 1991; 28:692–9North, RB Ewend, MG Lawton, MT Kidd, DH Piantadosi, S
North RM, Kidd DH, Farrokhi F, Piantadosi SA: Spinal cord stimulation versus  repeated lumbosacral spine surgery for chronic pain: A randomized, controlled trial. Neurosurgery 2005; 56:98–107North, RM Kidd, DH Farrokhi, F Piantadosi, SA
North RB, Kidd D, Shipley J, Taylor RS: Spinal cord stimulation versus  reoperation for failed back surgery syndrome: A cost effectiveness and cost utility analysis based on a randomized, controlled trial. Neurosurgery 2007; 61:361–9North, RB Kidd, D Shipley, J Taylor, RS
Honeys DD, Rashbaum RF, Bogdanffy GM: Prospective outcome evaluation of spinal cord stimulation I: Patients with intractable leg pain. Spine 1996; 21:1344–50Honeys, DD Rashbaum, RF Bogdanffy, GM
Ohnmeiss DD, Rashbaum RF: Patient satisfaction with spinal cord stimulation for predominant complaints of chronic, intractable low back pain. Spine J 2001; 1:358–63Ohnmeiss, DD Rashbaum, RF
Quigley DG, Arnold J, Eldridge PR, Cameron H, McIvor K, Miles JB, Varma TRK: Long-term outcome of spinal cord stimulation and hardware complications. Stereotact Funct Neurosurg 2003; 81:50–6Quigley, DG Arnold, J Eldridge, PR Cameron, H McIvor, K Miles, JB Varma, TRK
Rainov NG, Heidecke V, Burkert W: Short test-period spinal cord stimulation for failed back surgery syndrome. Minim Invasive Neurosurg 1996; 39:41–4Rainov, NG Heidecke, V Burkert, W
Sindou MP, Mertens P, Bendavid U, García-Larrea L, Mauguière F: Predictive value of somatosensory evoked potentials for long-lasting pain relief after spinal cord stimulation: Practical use for patient selection. Neurosurgery 2003; 52:1374–83Sindou, MP Mertens, P Bendavid, U García-Larrea, L Mauguière, F
Van Buyten JP, Van Zundert J, Vueghs P, Vanduffel L: Efficacy of spinal cord stimulation: 10 years of experience in a pain centre in Belgium. Eur J Pain 2001; 5:299–307Van Buyten, JP Van Zundert, J Vueghs, P Vanduffel, L
Villavicencio AT, Leveque J-C, Rubin L, Bulsara K, Gorecki JP: Laminectomy versus  percutaneous electrode placement for spinal cord stimulation. Neurosurgery 2000; 46:399–405Villavicencio, AT Leveque, J-C Rubin, L Bulsara, K Gorecki, JP
Juratli SM, Mirza SK, Fulton-Kehoe D, Wickizer TM, Franklin GM: Mortality after lumbar fusion surgery. Spine 2009; 34:740–7Juratli, SM Mirza, SK Fulton-Kehoe, D Wickizer, TM Franklin, GM
Oldridge MD, Yuan Z, Stoll JE, Rimm AR: Lumbar spine surgery and mortality among Medicare beneficiaries, 1986. Am J Public Health 1994; 84:1292–8Oldridge, MD Yuan, Z Stoll, JE Rimm, AR
Ramirez LF, Thisted R: Using a national health care data base to determine surgical complications in community hospitals: Lumbar discectomy as an example. Neurosurgery 1989; 25:218–25Ramirez, LF Thisted, R
Ramirez LF, Thisted R: Complications and demographic characteristics of patients undergoing lumbar discectomy in community hospitals. Neurosurgery 1989; 25:226–31Ramirez, LF Thisted, R
50. Rothman KJ, Boice JD: Epidemiologic Analysis with a Programmable Calculator (NIH Publication 79-1649). Washington DC, US Government Printing Office, 1979
51. Eckholm E, Pierce O: Methadone Rises as a Painkiller With Big Risks. NY Times August17, 2008
52. Meier B: FDA weighs training to dispense narcotics. NY Times August17, 2008
Phillips DP, Barker GE, Eguchi MM: A steep increase in domestic fatal medication errors with use of alcohol and/or street drugs. Arch Intern Med 2008; 168:1561–6Phillips, DP Barker, GE Eguchi, MM
Walker JM, Farney RJ, Rhondeau SM, Boyle KM, Cloward TV, Shilling KC: Chronic opioid use is a risk factor for the development of central sleep apnea and ataxic breathing. J Clin Sleep Med 2007; 3:455–61Walker, JM Farney, RJ Rhondeau, SM Boyle, KM Cloward, TV Shilling, KC
Bailey PL, Lu JK, Pace NL, Orr JA, White JL, Hamber EA, Slawson MH, Crouch DJ, Rollins DE: Effects of intrathecal morphine on the ventilatory response to hypoxia. N Engl J Med 2000; 343:1228–34Bailey, PL Lu, JK Pace, NL Orr, JA White, JL Hamber, EA Slawson, MH Crouch, DJ Rollins, DE
Hill CS Jr, The Texas Cancer Council's Workgroup on Pain Control in Cancer Patients: Guidelines For Treatment of Cancer Pain, 2nd edition. Austin, Texas Cancer Council, 1997, pp 25–6Hill, CS Austin Texas Cancer Council
Table 1. Information and Analyses for Each Data Source 
Image Not Available
Table 1. Information and Analyses for Each Data Source 
×
Table 2. Summary of Nine Index Cases of Death within 3 Days after Intrathecal Opioid Device Implant 
Image Not Available
Table 2. Summary of Nine Index Cases of Death within 3 Days after Intrathecal Opioid Device Implant 
×
Table 2. Continued 
Image Not Available
Table 2. Continued 
×
Table 3. Unadjusted Death Rates for Benchmark Therapies: Lumbar Diskectomy, Lumbosacral Spine Surgery, Spinal Cord Stimulation Implantation, and Intrathecal Drug Delivery Implantation for Noncancer Pain 
Image Not Available
Table 3. Unadjusted Death Rates for Benchmark Therapies: Lumbar Diskectomy, Lumbosacral Spine Surgery, Spinal Cord Stimulation Implantation, and Intrathecal Drug Delivery Implantation for Noncancer Pain 
×
Table 4. Deaths/Expected Deaths Ratios for Intrathecal Opioid and Spinal Cord Stimulation Patients by Time Period 
Image Not Available
Table 4. Deaths/Expected Deaths Ratios for Intrathecal Opioid and Spinal Cord Stimulation Patients by Time Period 
×
Table 5. Time of Death in Noncancer Pain Patients in Relation to a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
Image Not Available
Table 5. Time of Death in Noncancer Pain Patients in Relation to a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
×
Table 6. Reported Cause of Death in Noncancer Pain Patients within 3 Days of a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
Image Not Available
Table 6. Reported Cause of Death in Noncancer Pain Patients within 3 Days of a Pump Implant, Refill, Replacement, Programming, Dose Change, or Catheter Revision–From Enterprise Product Comment Report System 
×
Table 7. Noncancer Pain Patient Deaths within 3 Days by Pump- or Device-related Procedure–From Enterprise Product Comment Report System 
Image Not Available
Table 7. Noncancer Pain Patient Deaths within 3 Days by Pump- or Device-related Procedure–From Enterprise Product Comment Report System 
×