Background

The aseptic techniques of anesthesiologists in the preparation and administration of injected medications have not been extensively investigated, but emerging data demonstrate that inadvertent lapses in aseptic technique may be an important contributor to surgical site and other postoperative infections.

Methods

A prospective, open, microbiological audit of 303 cases in which anesthesiologists were asked to inject all bolus drugs, except propofol and antibiotics, through a 0.2-µm filter was performed. The authors cultured microorganisms, if present, from the 0.2-µm filter unit and from the residual contents of the syringes used for drawing up or administering drugs. Participating anesthesiologists rated ease of use of the filters after each case.

Results

Twenty-three anesthesiologists each anesthetized up to 25 adult patients. The authors isolated microorganisms from filter units in 19 (6.3%) of 300 cases (3 cases were excluded), including Staphylococcus capitis, Staphylococcus warneri, Staphylococcus epidermidis, Staphylococcus haemolyticus, Micrococcus luteus/lylae, Corynebacterium, and Bacillus species. The authors collected used syringes at the end of each case and grew microorganisms from residual drug in 55 of these 2,318 (2.4%) syringes including all the aforementioned microorganisms and also Kocuria kristinae, Staphylococcus aureus, and Staphylococcus hominus. Participants’ average rating of ease of use of the filter units was 3.5 out of 10 (0 being very easy and 10 being very difficult).

Conclusions

Microorganisms with the potential to cause infection are being injected (presumably inadvertently) into some patients during the administration of intravenous drugs by bolus during anesthesia. The relevance of this finding to postoperative infections warrants further investigation.

Anesthesiologists were asked to make bolus injections of all drugs, except propofol and antibiotics, through a 0.2-μm filter in a prospective, open, microbiological audit of 300 cases. Microorganisms with the potential to cause infections were isolated from the 0.2-μm filters of 19 (6.3%) of the 300 cases.

What We Already Know about This Topic
  • Postoperative infections represent a significant proportion of healthcare-associated infections

  • Anesthesiologists’ aseptic technique when making bolus injections of drugs may sometimes be deficient and could lead to postoperative infection

  • Anesthesiologists make an average of 10 bolus injections per case

What This Article Tells Us That Is New
  • Anesthesiologists were asked to make bolus injections of all drugs, except propofol and antibiotics, through a 0.2-μm filter in a prospective, open, microbiological audit of 300 cases

  • Microorganisms with the potential to cause infections were isolated from the 0.2-μm filters of 19 (6.3%) of the 300 cases

POSTOPERATIVE infections (e.g., surgical site infections, blood stream infections, and pneumonia) form a substantial proportion of healthcare-associated infections.1  The potential contribution of anesthesiologists to these infections has largely been overlooked, although it is clear that at least some of their practices could influence the rates at which they occur. For example, ensuring the completion of a bundle of care practices reduced the median rate of bloodstream infections associated with central line insertion from 2.7 per 1,000 catheter-days preintervention to 0 at 3 months,2  albeit in the context of intensive care. This suggests that other aspects of anesthesiologists’ practices could also (a) contribute to postoperative infections and (b) be amenable to improvement.

Anesthesiologists inject many intravenous drugs during every anesthetic—on average 10 bolus injections per patient and sometimes more than 100.3,4  In a busy perioperative environment, where multiple drugs are drawn up and injected, sometimes with urgency, it is not surprising that aseptic technique may occasionally lapse. The aseptic techniques of anesthesiologists in the preparation and administration of injected medications have not been extensively investigated, but emerging data5–11  demonstrate that inadequate aseptic technique may be an important, but previously overlooked, contributor to postoperative infections. In a recent study of 20 highly realistic simulated anesthetics using real drugs and standard practices, our group collected the injected drugs and associated intravenous fluid: we isolated Gram-positive staphylococci, Gram-negative bacilli, and other microorganisms from 5 of 38 bags (13%) of this collected injectate.5 

The clinical relevance of this disconcerting finding is not clear. In particular, it is difficult to be certain that this result was not an artifact of simulation: perhaps, the same anesthesiologists would have paid more attention to aseptic technique in a clinical setting.

Therefore, we aimed to ascertain the extent to which microorganisms are injected by anesthesiologists in the process of drawing up and administering boluses of intravenous drugs in clinical practice.

The clinical study was conducted in adult patients undergoing surgery with anesthesia in 19 operating rooms (ORs) at Auckland City Hospital, Auckland, New Zealand, a tertiary teaching hospital. The laboratory work was carried out in the Department of Molecular Medicine and Pathology, University of Auckland, Auckland, New Zealand. The clinical study was approved by the Northern B Health and Disabilities Ethics Committee, Ministry of Health, Wellington, New Zealand (ref: 13/NTB/80) and registered with the Australian New Zealand Clinical Trials Registry (ref: ACTRN 12613000040763) (Alan F. Merry, Principle investigator; date registered: January 14, 2013). The ethics committee determined that this study was, in effect, a low-risk audit of the participating anesthesiologists’ practice. Therefore, formal informed consent was not required from the patients. Instead, they were provided with an information sheet explaining the study, given the opportunity to discuss it with one of the research team, and given the opportunity to opt out of the study if they wished.

We tested the hypothesis that microbiological contamination of intravenous drugs administered during anesthesia occurs more often than 1 of 100 cases (a case being defined as one patient undergoing one anesthetic on one occasion3 ). To this end, we asked participating anesthesiologists to inject their intravenous bolus medications, except propofol (which could not be filtered because of its presentation in an emulsion), drugs given by infusion (because our primary interest was drugs given by intravenous bolus), and antibiotics (because they would inhibit the growth of contaminants) through a 0.2-µm filter unit (B. Braun Melsungen AG, Germany). We expected the filter membranes housed within the unit to trap any microorganisms inadvertently introduced during these injections. However, these units are sealed, which effectively prohibits extracting the membrane for culture without considerable risk of introducing contamination in the process. Therefore, we began by developing a method of quantifying the microorganisms trapped by the filter unit that we could apply to our clinical study.

A Method to Quantify Microorganisms Trapped on the Membrane of a Sealed Filter Unit

Informal experimentation demonstrated that flushing the B. Braun filter unit in the opposite direction to that used for injecting drugs (“backflushing”) with 20 ml of a combination of sterile 3% tryptic soy broth, 1% Tween 80, and peptone water was effective in allowing us to release and culture Staphylococcus epidermidis previously inoculated onto the filter membrane in loads of approximately 105 colony-forming units (CFUs) per milliliter.

To obtain more precise quantification of the effectiveness of this backflushing method, we next prepared standard dilutions of four microorganisms: S. epidermidis ATCC 12228, Escherichia coli ATCC 25922, Streptococcus pyogenes ATCC 8668, and Candida albicans ATCC 1212. We cultured these microorganisms at 37°C on horse blood agar or in tryptic soy broth (both Fort Richard, New Zealand). Working with one organism at a time, we harvested an overnight culture by centrifugation at 4,000g at 22°C for 10 min and resuspended this in 0.9% sodium chloride (Demo S.A., Greece) at approximately 108 CFU/ml. We made serial dilutions from this suspension under aseptic conditions producing four solutions with four levels of contamination and a control of sterile 0.9% sodium chloride. We injected 10 ml of each of these five solutions through a filter unit. The five filter units were kept under aseptic conditions at 22°C for 1 h to reflect the shortest time interval taken in the clinical study to retrieve filter units from the OR and return with them to the microbiology laboratory for further testing.

We then backflushed 20 ml of combined sterile 3% tryptic soy broth, 1% Tween 80, and peptone water through each of the five filter units using a pulsatile technique.12  The backflushed fluid was collected in five separate, sterile 20-ml syringes (Becton, Dickinson and Company, USA), and then the content of each syringe was injected through another sterile but differently designed filter unit containing a 0.2-µm membrane. These secondary filter units are not approved for clinical use but are used in the laboratory where they are typically repeatedly resterilized and reused. They can be dismantled, which makes it possible to extract the membrane easily (both unit and membrane from Pall Corporation, USA). The five membranes from these units were retrieved and placed onto a separate blood agar plate. These were incubated at 37°C for 24 to 72 h, and the resulting CFU count was used as a measure of the microorganisms released by backflushing. We also collected the fluid passing through the secondary filter units during this second filtration and incubated it at 37°C for 24 h to establish whether all microorganisms had been captured by the membrane. We repeated this entire sequence five times with each of the five solutions for each of the four microorganisms.

The Clinical Investigation

We approached anesthesiologists at levels of seniority ranging from junior registrar to senior consultant working within the anesthesia service at Auckland City Hospital and obtained written informed consent from those who agreed to participate. Cases were eligible if scheduled for general anesthesia (with or without regional anesthesia) and cared for by participating anesthesiologists on days of the week when one of the investigators (Derryn Gargiulo) was available.

A filter unit and stopcock were incorporated into the drug administration line via a two-way extension set (both CareFusion, USA) by the participant or anesthetic technician after instruction by Derryn Gargiulo. The volume of dead space in the stopcock, filter unit, and filter arm of the two-way extension line (the filter unit “setup”) was 1.16 ml (fig. 1). Therefore, the participating anesthesiologists were asked to flush the setup with 0.9% sodium chloride (Demo S.A.; drawn up by them) each time a drug was administered through the setup.

Fig. 1.

Clinical setup for the injection port (i), filter unit (ii), and two-way extension line (iii).

Fig. 1.

Clinical setup for the injection port (i), filter unit (ii), and two-way extension line (iii).

Close modal

A label was attached to the filter arm of the two-way extension line indicating that it was for medications only, except propofol or antibiotics (fig. 2).

Fig. 2.

Label attached to the filter arm of the two-way extension line.

Fig. 2.

Label attached to the filter arm of the two-way extension line.

Close modal

Intravenous fluids were administered via the nonfilter arm of the two-way extension. Propofol, drug infusions, and antibiotics were administered via a secondary stopcock available for this purpose on the fluid line, thus bypassing the filter unit. Participants were able to use one or more additional intravenous lines and cannulae for additional intravenous fluids or infusions at their discretion.

At the end of each case, the filter unit was removed, and its two ends were sealed with sterile caps (fig. 1). They were taken to the laboratory in zip-locked plastic bags for backflushing and subsequent analysis as described in A Method to Quantify Microorganisms Trapped on the Membrane of a Sealed Filter Unit section. Syringes used to prepare and administer medications during each case were sealed with sterile caps, taken to the laboratory and residual contents analyzed for contamination using quantitative methods described in our previous study.5 

Care was taken to ensure that no microorganisms were introduced during the handling of the filter units and the syringes. In the laboratory, clean gowns and hats were worn, and sterile gloves were used. All work was undertaken in a validated laminar flow cabinet. Sterile 70% alcohol was used to decontaminate all work surfaces and the external surfaces of the filter units and syringes.13,14  Senior microbiologists provided oversight of the processes of handling the filter units and syringes in the OR and the laboratory. Two independent senior anesthesiologists acted as monitors of our study method in the OR.

To evaluate the potential for false positives from contamination by the investigators during the testing of the filter units, the backflushing method and analysis were audited with 10 filter units analyzed on each of four separate occasions in conjunction with filter units collected from cases. We injected 10 ml of control solution through these audit filter units and then processed them in the same way as the patient’s filter units.

If any microorganisms were grown from membranes or syringe contents, the agar plates were sent to the Clinical Microbiology Laboratory, LabPlus, Auckland City Hospital for identification of the organisms using the matrix-assisted laser desorption/ionization time of flight technique.15 

At the end of each case, participating anesthesiologists were asked to complete a short form confirming that the study protocol had been followed or giving reasons if not (i.e., if the filter unit setup had been bypassed). They were also asked to rate the ease of use of the setup using an unmarked 10-cm visual analog scale (created for this study) with anchors “very easy” (at 0 cm) and “very difficult” (at 10 cm) and to provide any comments if they so wished. The mean difficulty rating was calculated for each anesthesiologist, and then the mean of these means was calculated.

Sample Size

Our sample size estimate was pragmatic. In our previous simulation-based study,5  contamination was identified in 13% of cases (5 of 38 bags of collected injectate and intravenous fluid grew microorganisms). We thought that a rate of at least one tenth of this would be very likely in the clinical setting—in other words, more than 1%. We thought, furthermore, that a rate of more than or equal to 1% would be clinically important, even allowing for the fact that only a proportion of injected bacteria are likely to cause harm. Taking these factors into account, we predicated our sample size on the potential interpretation of a negative result. In the event that no bacteria were cultured in any of our cases, it would require 300 cases to ensure that the upper 95% CI of the estimated point incidence of 0% was no higher than 1% and, therefore, allow us to reject our hypothesis with an α of 0.05.

Statistics

Data were entered into a spreadsheet (Microsoft Corporation, USA, 2010), and summary statistics were calculated.

Laboratory Testing

Our method of diluting the broth for preloading the filters produced a range of concentrations of microorganisms in our inoculants. We were able to quantify microorganisms approximately, but reasonably consistently, from filter units loaded from these inoculants, even when the concentrations of microorganisms (estimated from plate counts) were less than 10 CFU. Below this value, the results were less reliable, but organisms could still be grown in most instances.

The fluid collected from the second filtration by the secondary filter unit produced no growth (as shown by clear solutions after incubation), indicating that all microorganisms present in the filtrate were captured by the membrane within the second filter unit.

The quality assurance testing of the “retrieval” method using 10 ml of control solution as inoculant was carried out on 38 of the 40 audit filter units because on two occasions, a sterile secondary filter unit was not available. No growth was obtained from any of these 38 units.

The Clinical Study

Patients, Cases, and Participants.

Data were collected from 303 cases over 76 days between August 13, 2013, and June 16, 2014. Three cases were excluded because, respectively, one filter unit was lost in transit, one filter unit broke during testing in the laboratory, and there was a failure in timely supply of the testing supplies on one occasion; thus, the final analysis of filter units was undertaken on 300 cases from 298 patients (2 patients each had a second, different operation).

The cases were managed by 23 participating anesthesiologists (19 specialists, 2 registrars, and 2 fellows) of whom 11 were women and 20 were right handed. Participants managed a mean (range) of 13 (1 to 25) cases each. The distribution of cases with respect to order of the day, day of week, surgical specialty, and duration of anesthesia is presented in table 1.

Table 1.

The Order in the Day, Day of Week, Surgical Specialty, and Duration of Anesthesia of the Cases in the Study

The Order in the Day, Day of Week, Surgical Specialty, and Duration of Anesthesia of the Cases in the Study
The Order in the Day, Day of Week, Surgical Specialty, and Duration of Anesthesia of the Cases in the Study

Microbiology.

In total, 301 filter units from 300 cases were tested (two filter units were used on two different intravenous lines in one cardiac case at the request of the participant anesthesiologist). Microorganisms were grown from 19 (6.3%; 95% CI, 4.1% to 9.7%) of these filter units (table 2), collected from 19 cases. These cases were managed by 12 (50%) of the participants: 2 participants managed three cases, 3 more participants managed two cases, and a further 7 participants each managed one case (table 2).

Table 2.

Number of Cases Undertaken, Number of Syringes Collected from Each Participant, Number of Cases with Contaminated Filter Units, and Number of Syringe Contents Contaminated

Number of Cases Undertaken, Number of Syringes Collected from Each Participant, Number of Cases with Contaminated Filter Units, and Number of Syringe Contents Contaminated
Number of Cases Undertaken, Number of Syringes Collected from Each Participant, Number of Cases with Contaminated Filter Units, and Number of Syringe Contents Contaminated

In total, 2,318 syringes were collected from 295 of the cases, a mean (range) of 7 (0 to 16) syringes per case. The remaining five cases were excluded because there was a noted accident in maintaining the integrity of the aseptic collection of the syringes in two cases, one case was interrupted and then abandoned (but the filter unit still collected for analysis), and the syringes were accidently discarded by OR staff in two cases. In 48 (16%) cases managed by 18 (78%) participants, microorganisms were grown and identified from the contents of 55 syringes (tables 2 and 3).

Table 3.

Microorganisms Recovered from Filter Membranes and Syringe Contents, by Participant and Case

Microorganisms Recovered from Filter Membranes and Syringe Contents, by Participant and Case
Microorganisms Recovered from Filter Membranes and Syringe Contents, by Participant and Case

Growth was obtained from both the filter units and (some of the) syringes used by 11 (48%) participants; from the syringes, but not the membranes used by 7 (30%) participants; and from neither the membranes nor the syringes used by 4 (18%) participants. There was one participant from whom we obtained growth from each filter unit in three cases, but no growth in any of the syringes (table 3, participant 3). There was only one case in which the same organism was grown from a filter unit and a syringe (table 3, participant 1).

The cases were managed in 19 ORs. The mean (range) of cases per room was 16 (1 to 52). The cases in which growth was obtained from filters came from 12 ORs and those with growth from syringes came from 16 ORs; 11 ORs had cases where growth was obtained from both filter units and syringes.

Participants’ Feedback.

In five cases, the participant indicated that propofol was inadvertently injected through the filter unit, and that in two cases, an antibiotic was injected (no microorganisms were grown from the filter unit in any of these seven cases). In 96 cases, the filter unit was bypassed at least once (table 4).

Table 4.

Reasons for Bypassing Filter Unit (More than One Reason Could Be Given for Each Case)

Reasons for Bypassing Filter Unit (More than One Reason Could Be Given for Each Case)
Reasons for Bypassing Filter Unit (More than One Reason Could Be Given for Each Case)

The 23 participating anesthesiologists’ mean ratings of ease of use of the filter units ranged from 0.1 to 7.2, with an overall mean of 3.5. In this analysis, 4 of the 300 cases were excluded because the responses were not clear, and 1 case was excluded because no response was given.

The participants’ additional comments regarding the study included (among others) references to the extra pressure required to inject drugs though the filter units (11 cases); the consequent slowness of the injections (3 cases); and changes to normal behavior (10 cases).

We trapped and grew potentially pathogenic microorganisms injected intravenously during the bolus administration of intraoperative drugs in 6.3% of 300 cases in which patients underwent general anesthesia. We have no reason to doubt the sterility of the drugs provided in ampules or vials and no reason to doubt that their subsequent contamination was other than inadvertent. In 16% of cases, we grew microorganisms from the residual contents of syringes that had been retained by the anesthesiologists, and which might potentially have been used again for further intravenous injections. Collectively, the varieties of microorganism grown from the filter units and syringes were similar, but we found little direct concordance in individual cases between microorganism isolates from the filter units through which drugs were injected and residual drug in retained syringes that had been used to inject these drugs (tables 2 and 3). Some of this species variation may be due to the matrix-assisted laser desorption/ionization time of flight identification: there may be less variation in deoxyribonucleic acid and/or bacterial resistance profile. The precise sources of contamination, and the aspects of practice that need to be addressed to prevent contamination, cannot be determined from this study. Nevertheless, these findings corroborate our previous data5  and that of others,6–11  which suggest that anesthesiologists’ aseptic techniques, in relation to the injection of drugs by bolus, may sometimes be deficient.

To what extent were the isolates retrieved from the filter units a true reflection of our patients’ actual exposure to microorganisms injected inadvertently with intravenous bolus drugs during anesthesia? It is difficult to completely rule out the possibility that at least some of the cultured microorganisms could have been introduced during the collection and flushing of the filter units. However, we took considerable care to avoid introducing microorganisms in our handling of the filters and syringes. Furthermore, the data from the testing of the 38 sterile filter units included to audit this aspect of the study are reassuring.

One related limitation in our development of the backflush technique is that only four microorganisms were tested. These were chosen to align with previous research,5  to include common environmental16  and population-based microorganisms17  likely to be associated with postoperative infections, and to include an example of a coccus, a bacillus, and a yeast. Other microorganisms, e.g., endospore-forming bacteria may behave differently. Similarly, only one type of commercial filter unit was studied; alternative brands with different filter materials may show different characteristics when backflushed.

Weaknesses of this study include the use of a sample of convenience and the fact that the study was conducted in one center only; these results might not apply to other anesthesiologists or other centers. On the other hand, there is no particular reason to assume that that the practices evaluated here would be in any way unusual. We acknowledge that relieving anesthesiologists, who were not participants, managed the participants’ cases for periods of time, but we do not believe that this changes the clinical relevance of our finding—our study was not directed at individual practitioners, but rather at the overall process by which intravenous drugs are administered to patients during anesthesia. Our rating scale for ease of use of the filters was not formally validated, but it was similar to visual analog scales used in many previous studies (e.g., in the study by Merry et al.3 ). We assumed that the 0.2-µm filter units performed to specifications, but we did not verify this; the filter unit was bypassed at least once in almost a third of our cases for various reasons (table 4): these points also have implications for the potential utility of such filters in addressing the problem of inadvertently injected microorganisms.

The potential confounding of our data by failure of the filter units to trap all organisms, or failure of the backflush method to retrieve all trapped microorganisms, could have resulted in our underestimating the exposure of patients to microorganisms by the intravenous route. In addition, although anesthesiologists were encouraged to behave “normally” in respect of their aseptic practice, the open-label nature of the study may have influenced them to be more fastidious. Given propofol’s known ability to promote the growth of microorganisms,18  its exclusion from injection through the filter units is also relevant. Thus, there were several potential confounders that could have resulted in our underestimating of the rate of intravenous injection of microorganisms. However, the use of the filter unit on a Y-connector and the associated need to flush boluses of drug into the intravenous line using sodium chloride involved more opportunities for contamination. Any failures in our own handling of filter units and syringes might also have inflated our results. Therefore, it is possible that our data could either overestimate or underestimate the true incidence of intravenous injection of microorganisms. Ultimately, it is the order of magnitude of our result that matters, more than the exact rate. We selected 1% as the threshold for clinical concern in our hypothesis; this was a subjective judgment, which took into account factors discussed in the following paragraph, but, arguably, the injection of intravenous drugs should be accomplished in a sterile fashion, and in the context of processes control, it would be reasonable to expect a failure rate very close to 0 and certainly less than 1%.

The extent to which microorganisms from the bolus injection of intravenous drugs might contribute to postoperative infections is not clear. Microorganisms may be present in the blood stream without causing harm, after brushing one’s teeth for example.19  In at least some patients, any injected microorganisms may be adequately dealt with by the immune system and through the use of prophylactic antibiotics (which are given routinely for many surgical procedures). However, the operative wound is an ideal environment for microorganisms, seeded through the bloodstream, to establish infection, particularly given that many patients undergoing surgery and anesthesia are debilitated or ill or may have reduced immune responses.20–22  It seems at least plausible that injecting microorganisms in this way could contribute to some postoperative infections in at least some patients. Recent research has found 18.6% of injection ports (on stopcocks) in intravenous lines used for drugs other than propofol and 17.3% of those used for propofol to be contaminated with microorganisms at 6 h after first use.23  This potential source of contamination may be additional to that demonstrated by our results or may explain some or all of our results. A filter strategy would likely be an effective way to reduce blood stream contamination in either case, at least for drugs other than propofol, because the injection port on the filter unit is proximal to the filter membrane. In the end, the clinical relevance of these potential sources of infection will need to be evaluated through a randomized controlled trial of an intervention to prevent or at least reduce them.

The question arises, then, of how one might reduce the frequency with which anesthesiologists inadvertently inject microorganisms while injecting intravenous drugs through injection ports, given that it is difficult to change embedded clinical practices (this can be seen, e.g., in the difficulty improving practice in relation to hand hygiene24–26 ). Our ease-of-use ratings confirm that it would be practical for anesthesiologists to routinely include 0.2-µm filter units into their cases. It would not be possible to inject propofol through the filter units, and, as discussed earlier, some microorganisms might pass through the filters, but the majority of bolus injections could be filtered, and the load of injected microorganisms could be substantially reduced. Therefore, we plan a study in which we will test the hypothesis that using filter units of this type will reduce subsequent postoperative infections. Our results reinforce the importance of meticulous aseptic technique in administering intravenous injections, particularly when using high stakes access points, such as central venous catheters and peripherally inserted central catheters (where the opportunity for catheter-related blood stream infection is ever present). The routine use of an alcohol wipe of the septum before accessing drug vials may also warrant emphasis (data from the study by Hilliard et al.27  and from our previous simulation-based study5  suggest that the value of this may not be fully appreciated in New Zealand and elsewhere).

In the meantime, we conclude that microorganisms with the potential to cause infection are being injected into at least some patients during the administration of intravenous bolus drugs during anesthesia. Clearly this could include any pathogen present in the OR environment.28  Strategies to reduce this potential source of infection should be developed.

The authors thank the anesthesiologists, anesthetic technicians, and patients who participated in this study. They also thank the operating room and PACU staff on levels 4 and 8 at Auckland City Hospital, Auckland, New Zealand, for their support. They gratefully acknowledge the assistance of Mary Bilkey, B.Sc., Microbiology Department, LabPlus, Auckland, New Zealand; Sally Roberts, M.B.Ch.B., Ph.D., Clinical Head of Microbiology, LabPlus; Charles Bradfield, M.B.B.Ch., F.A.N.Z.C.A., Level 8 Anaesthesia and Operating Rooms, Auckland City Hospital; and Nelis Kruger, M.B.Ch.B.(Pret), F.A.N.Z.C.A., Cardiothoracic and ORL Anaesthesia, Auckland City Hospital, for conducting the microbiological and clinical audits. They also thank Anne Familton, B.Pharm., from Becton Dickinson (Auckland, New Zealand) for the supplies of the PosiFlush 0.9% sodium chloride syringes; B. Braun (Auckland, New Zealand) for the procurement of the 0.2-μm filter units; and the staff and fellow students in Molecular Medicine and Pathology, the University of Auckland, Auckland, New Zealand, for their microbiological advice. They also thank the technical staff at the School of Pharmacy, Auckland, New Zealand, for their assistance, and the Kate Edger Educational Charitable Trust (Auckland, New Zealand) for providing funds for Derryn Gargiulo to allow her to complete the data collection free from teaching duties.

This study was supported by a doctoral completion award from the Kate Edger Educational Charitable Trust and by the University of Auckland as part of doctoral research (to Derryn Gargiulo).

The authors declare no competing interests.

Full protocol available from Dr. Merry: a.merry@auckland.ac.nz. Raw data available from Dr. Merry: a.merry@auckland.ac.nz.

1.
World Health Organization
:
The Burden of Endemic Health Care-associated Infection Worldwide
.
Geneva
,
World Health Organization
,
2011
Available at: http://www.who.int/gpsc/country_work/burden_hcai/en/. Accessed June 9 2015
2.
Pronovost
P
,
Needham
D
,
Berenholtz
S
,
Sinopoli
D
,
Chu
H
,
Cosgrove
S
,
Sexton
B
,
Hyzy
R
,
Welsh
R
,
Roth
G
,
Bander
J
,
Kepros
J
,
Goeschel
C
:
An intervention to decrease catheter-related bloodstream infections in the ICU.
N Engl J Med
2006
;
355
:
2725
32
3.
Merry
AF
,
Webster
CS
,
Hannam
J
,
Mitchell
SJ
,
Henderson
R
,
Reid
P
,
Edwards
KE
,
Jardim
A
,
Pak
N
,
Cooper
J
,
Hopley
L
,
Frampton
C
,
Short
TG
:
Multimodal system designed to reduce errors in recording and administration of drugs in anaesthesia: Prospective randomised clinical evaluation.
BMJ
2011
;
343
:
d5543
4.
Webster
CS
,
Merry
AF
,
Gander
PH
,
Mann
NK
:
A prospective, randomised clinical evaluation of a new safety-orientated injectable drug administration system in comparison with conventional methods.
Anaesthesia
2004
;
59
:
80
7
5.
Gargiulo
DA
,
Sheridan
J
,
Webster
CS
,
Swift
S
,
Torrie
J
,
Weller
J
,
Henderson
K
,
Hannam
J
,
Merry
AF
:
Anaesthetic drug administration as a potential contributor to healthcare-associated infections: A prospective simulation-based evaluation of aseptic techniques in the administration of anaesthetic drugs.
BMJ Qual Saf
2012
;
21
:
826
34
6.
Loftus
RW
,
Koff
MD
,
Burchman
CC
,
Schwartzman
JD
,
Thorum
V
,
Read
ME
,
Wood
TA
,
Beach
ML
:
Transmission of pathogenic bacterial organisms in the anesthesia work area.
Anesthesiology
2008
;
109
:
399
407
7.
Loftus
RW
,
Muffly
MK
,
Brown
JR
,
Beach
ML
,
Koff
MD
,
Corwin
HL
,
Surgenor
SD
,
Kirkland
KB
,
Yeager
MP
:
Hand contamination of anesthesia providers is an important risk factor for intraoperative bacterial transmission.
Anesth Analg
2011
;
112
:
98
105
8.
Loftus
RW
,
Patel
HM
,
Huysman
BC
,
Kispert
DP
,
Koff
MD
,
Gallagher
JD
,
Jensen
JT
,
Rowlands
J
,
Reddy
S
,
Dodds
TM
,
Yeager
MP
,
Ruoff
KL
,
Surgenor
SD
,
Brown
JR
:
Prevention of intravenous bacterial injection from health care provider hands: The importance of catheter design and handling.
Anesth Analg
2012
;
115
:
1109
19
9.
Loftus
RW
,
Brown
JR
,
Koff
MD
,
Reddy
S
,
Heard
SO
,
Patel
HM
,
Fernandez
PG
,
Beach
ML
,
Corwin
HL
,
Jensen
JT
,
Kispert
D
,
Huysman
B
,
Dodds
TM
,
Ruoff
KL
,
Yeager
MP
:
Multiple reservoirs contribute to intraoperative bacterial transmission.
Anesth Analg
2012
;
114
:
1236
48
10.
Loftus
RW
,
Koff
MD
,
Brown
JR
,
Patel
HM
,
Jensen
JT
,
Reddy
S
,
Ruoff
KL
,
Heard
SO
,
Yeager
MP
,
Dodds
TM
:
The epidemiology of Staphylococcus aureus transmission in the anesthesia work area.
Anesth Analg
2015
;
120
:
807
18
11.
Loftus
RW
,
Koff
MD
,
Brown
JR
,
Patel
HM
,
Jensen
JT
,
Reddy
S
,
Ruoff
KL
,
Heard
SO
,
Yeager
MP
,
Dodds
TM
:
The dynamics of Enterococcus transmission from bacterial reservoirs commonly encountered by anesthesia providers.
Anesth Analg
2015
;
120
:
827
36
12.
Hadaway
L
:
Heparin locking for central venous catheters.
J Vasc Access
2006
;
11
:
224
31
13.
Anonymous
:
United States pharmacopeial convention: pharmaceutical compounding—sterile preparations
in
The United States Pharmacopeia
, 34th edition.
Rockville, MD
,
The United States Pharmacopeial Convention
,
2010
, pp
336
74
14.
PIC/S Secretariat
:
Guide to Good Manufacturing Practices for the Preparation of Medicinal Products
2014
Available at: www.picscheme.org. Accessed May 27, 2015
15.
Seng
P
,
Rolain
JM
,
Fournier
PE
,
La Scola
B
,
Drancourt
M
,
Raoult
D
:
MALDI-TOF-mass spectrometry applications in clinical microbiology.
Future Microbiol
2010
;
5
:
1733
54
16.
Edmiston
CE
Jr
,
Sinski
S
,
Seabrook
GR
,
Simons
D
,
Goheen
MP
:
Airborne particulates in the OR environment.
AORN J
1999
;
69
:
1169
72
17.
Huttenhower
C
,
Gevers
D
,
Knight
R
,
Abubucker
S
,
Badger
JH
,
Chinwalla
AT
,
Creasy
HH
,
Earl
AM
,
FitzGerald
MG
,
Fulton
RS
,
Giglio
MG
,
Hallsworth-Pepin
K
,
Lobos
EA
,
Madupu
R
,
Magrini
V
,
Martin
JC
,
Mitreva
M
,
Muzny
DM
,
Sodergren
EJ
,
Versalovic
J
:
Structure, function and diversity of the healthy human microbiome.
Nature
2012
;
486
:
207
14
18.
Bennett
SN
,
McNeil
MM
,
Bland
LA
,
Arduino
MJ
,
Villarino
ME
,
Perrotta
DM
,
Burwen
DR
,
Welbel
SF
,
Pegues
DA
,
Stroud
L
:
Postoperative infections traced to contamination of an intravenous anesthetic, propofol.
N Engl J Med
1995
;
333
:
147
54
19.
Silver
JG
,
Martin
AW
,
McBride
BC
:
Experimental transient bacteraemias in human subjects with clinically healthy gingivae.
J Clin Periodontol
1979
;
6
:
33
6
20.
Pawelec
G
:
Immunity and ageing in man.
Exp Gerontol
2006
;
41
:
1239
42
21.
Sanders
RD
:
III. Perioperative immunity: Is there an anaesthetic hangover?
Br J Anaesth
2014
;
112
:
210
2
22.
Gutierrez
T
,
Hornigold
R
,
Pearce
A
:
The systemic response to surgery.
Surgery
2014
;
32
:
149
52
23.
Cole
DC
,
Baslanti
TO
,
Gravenstein
NL
,
Gravenstein
N
:
Leaving more than your fingerprint on the intravenous line: A prospective study on propofol anesthesia and implications of stopcock contamination.
Anesth Analg
2015
;
120
:
861
7
24.
Allegranzi
B
,
Sax
H
,
Pittet
D
:
Hand hygiene and healthcare system change within multi-modal promotion: A narrative review.
J Hosp Infect
2013
;
83
(
suppl 1
):
S3
10
25.
Fernandez
PG
,
Loftus
RW
,
Dodds
TM
,
Koff
MD
,
Reddy
S
,
Heard
SO
,
Beach
ML
,
Yeager
MP
,
Brown
JR
:
Hand hygiene knowledge and perceptions among anesthesia providers.
Anesth Analg
2015
;
120
:
837
43
26.
Allegranzi
B
,
Storr
J
,
Dziekan
G
,
Leotsakos
A
,
Donaldson
L
,
Pittet
D
:
The First Global Patient Safety Challenge “Clean Care is Safer Care”: From launch to current progress and achievements.
J Hosp Infect
2007
;
65
(
suppl 2
):
115
23
27.
Hilliard
JG
,
Cambronne
ED
,
Kirsch
JR
,
Aziz
MF
:
Barrier protection capacity of flip-top pharmaceutical vials.
J Clin Anesth
2013
;
25
:
177
80
28.
Loftus
RW
,
Brown
JR
,
Patel
HM
,
Koff
MD
,
Jensen
JT
,
Reddy
S
,
Ruoff
KL
,
Heard
SO
,
Dodds
TM
,
Beach
ML
,
Yeager
MP
:
Transmission dynamics of gram-negative bacterial pathogens in the anesthesia work area.
Anesth Analg
2015
;
120
:
819
26